beautypg.com

Metrohm 785 DMP Titrino User Manual

Page 34

background image

2.6 Parameters, key

785 DMP Titrino

31

Relation between the stop criteria "time" and "drift"

The stop criterion "time", t(delay), means that the end point must be exceeded for a

certain period of time. In other words, after the last dosed increment, time t is allowed to

elapse before the titration is stopped. The size of this last increment depends on the

volume of the Exchange Unit used. With a 20 mL Exchange Unit, the smallest possible

increment is 2 ul. With a t(delay) = 5 s, the last 2 ul reagent must thus suffice for 5 s or

longer. This results in a drift of

2 uL/5 s = 24 uL/min (the drift can be less than 24

uL/min as it is not known whether the last increment would also suffice for 10 s). If you

have been working up to this point with a 20 mL Exchange Unit and a t(delay) = 5 s, you

can set a value of

24 uL/min as stop drift. The following table shows several values for

the maximum stop drift.

t(delay)

min.incr. (Exchange Unit)

5 s

10 s

20 s

0.5 uL (5 ml)

6 uL/min

3 uL/min

1.5 uL/min

1 uL (10 ml)

12 uL/min

6 uL/min

3 uL/min

2 uL (20 ml)

24 uL/min

12 uL/min

6 uL/min

5 uL (50 ml)

60 uL/min

30 uL/min

15 uL/min

Same t(delay) with a range of extremely small volume increments means different switch-

off points. In case the stop criterion "drift" is used, the stopping point remains the same.