2.6. Parameters, key

785 DMP Titrino

16

pause 0 s
**titr.

meas.input: 1

I(pol) 1 uA

U(pol) 400 mV

electrode test: OFF

temperature 25.0 °C

Waiting time (0...999 999 s)

Waiting time, e.g. for equilibration of the electrode after

the start or reaction time after dosing of start volume.

The pause can be aborted with .

Measuring input for pH and U (1, 2, diff.)

Request for measuring input for pH and U.

Measuring input 1 or 2 or differential amplifier;

connection of electrodes, see page 158.

With polarized electrodes, instead of the measuring

input the

polarization current (-127...127 uA)

or the

polarization voltage (-1270...1270 mV, in steps of 10

mV)

is inquired.

Electrode test (OFF, ON)

Test for polarized electrodes. Performed on

changeover from the inactive standby state to a

measurement. "OFF" means that the test is not

performed.

Titration temperature (-170.0...500.0 °C)

The temperature is continuously measured if a T

sensor is connected. This parameter is used for

temperature compensation in pH titrations.

>stop conditions

stop V: abs.
**titr.

stop V 99.99 ml
**titr.

factor 999999
**titr.

stop pH OFF
**titr.

Stop conditions for the titration

If several stop conditions have been set, the criterion

which is met first applies.

Type of stop volume (abs., rel., OFF)

"abs.": absolute stop volume in mL.

"rel.":

relative stop volume to sample size.

"OFF": stop volume switched off. Stop volume is not

monitored.

If "abs." is set:

Absolute stop volume (0...9999.99 mL)

If "rel." is set:

Factor for relative stop volume (0...

±999 999)

Calculated as: Stop V in mL = factor * sample size

Stop at measured value (input range dependent on

measured variable:

pH:

0...

± 20.00, OFF

U, Ipol: 0...

± 2000 mV, OFF

Upol:

0...

± 200.0 uA, OFF)

sets "OFF".

"OFF" means that the criterion is not monitored.