3.2. Remote control commands

785 DMP Titrino

114

Calibration data can be entered. They are accepted with &Info.CalibrationData
$G. If calibration data are entered, the calibration date is deleted.

3.2.2.94. Info.Checksums

$G

Info.Checksums.MPList

read only

Info.Checksums.ActualMethod

read only

The checksums can be used to identify the content of a file unequivocally,
e.g. files with identical content
have identical results of the checksums. An empty file has checksum "0". The
calculation of the checksums is triggered with $G.
.MPList:

Result of the checksum of the current measuring point list.

.ActualMethod: Result of the checksum of the current method in the working

memory. Identical methods with different method names have the
same results of the checksum.

3.2.2.95. Info.DetermData

$G

Info.DetermData.Write

ON, OFF

Info.DetermData.MPList.1.Attribute

read only/read+write

Info.DetermData.ExV

read only/read+write

Info.DetermData.MPList.1.X

read only/read+write

Info.DetermData.MPList.1.Y

read only/read+write

Info.DetermData.MPList.1.Z1

read only/read+write

Info.DetermData.MPList.1.Z2

read only/read+write

for every measuring point

Determination data in hexadecimal format. A measuring point list is available
in mode DET, MET, SET, and MEAS.
Recalculation of the measuring data is triggered with $G.
.Write:

With "ON", the following nodes can be overwritten:
&Info.DetermData.MP.List, &Info.TitrResults.Var.C4X (X = 0...5),
&Info.TitrResults.TempVar.C7X (X = 0...9), and &Mode.Name.

.ExV:

Volume of the exchange unit, with which the determination was
executed

.MPList.1.Attribute: Attribute
.MPList.X: X coordinate, time
.MPList.Y: Y coordinate, volume
.MPList.Z1: Z1 coordinate, measuring value
.MPList.Z2: Z2 coordinate, temperature

3.2.2.96. Info.TitrResults.RS.1.Value

read only

etc., up to .9
Info.TitrResults.EP.1.V

read only

Info.TitrResults.EP.1.Meas

read only

etc., up to .2
Info.TitrResults.Var.C40

read only/read+write

etc., up to .C47
Info.TitrResults.Var.DTime

read only/read+write

.RS: Values of the calculated results.
.EP: Endpoints with DET, MET, SET:

Volume coordinate in mL, e.g. "1.2340"