Metrohm 785 DMP Titrino User Manual
Page 100
3.2 Remote control commands
785 DMP Titrino
97
3.2.2 Description of the remote control commands
3.2.2.1.
Mode
$G, $S, $H, $C
Start and stop ($G, $S) or hold of the current method (3.2.2.3) with $H and
continue with $C.
$G also serves to continue after inquiries of identifications and sample size
after the start (see 3.2.2.25) as well as after inquiries of calibration tempera-
ture and pH values of buffers (see 3.2.2.41 and 3.2.2.42).
3.2.2.2.
Mode.QuickMeas
$G, $S
Start and stop of a measurement in the basic mode with the parameters
(measured quantity, measuring input) of the current method. Corresponds to
the
&Mode.Parameter.Presel, see 3.2.2.49.
With an ongoing measurement, the current mode can be started. This stops
the measurement automatically.
3.2.2.3.
Mode.Select
DET, MET, SET, MEAS, CAL, TIP
Mode.DETQuantity
pH, U, Ipol, Upol
Mode.METQuantity
pH, U, Ipol, Upol
Mode.SETQuantity
pH, U, Ipol, Upol
Mode.MEASQuantity
pH, U, Ipol, Upol, T
Selection of the standard mode. Mode and the measured quantity belong to
the complete selection.
If a method is selected from the method memory, the nodes &Mode.Select
and &Mode.XXXQuantity are overwritten with mode and measured quantity of
the corresponding user method.
3.2.2.4.
Mode.Name
read only
Name of the current method in the working memory. $Q sends 8 ASCII char-
acters. Standard methods carry the name ********. The node can be set
read + write, see 3.2.2.95.
3.2.2.5.
Mode.Def.Formulas.1.Formula
EPX, CXX, RSX, +, -, *, /, (, )
Mode.Def.Formulas.1.TextRS
up to 8 ASCII characters
Mode.Def.Formulas.1.Decimal
0...2...5
Mode.Def.Formulas.1.Unit
up to 6 ASCII characters
Mode.Def.Formulas.1.Limits
ON, OFF
Mode.Def.Formulas.1.LoLim
0...
±
999 999
Mode.Def.Formulas.1.UpLim
0...
±
999 999
Mode.Def.Formulas.1.Output
active, pulse, OFF
Mode.Def.Formulas.2.Formula
etc. up to .9
Entry of formulas. Rules for formula entry, see page 38.
Example: "(EP2-EP1)*C01/C00"
In addition to the formula, a text for result output, the number of decimal
places and a unit for the result output can be selected. "No unit" is selected
with the blank string.