4.2. Diagnosis

785 DMP Titrino

146

9 Memory card test

Memory card test

Memory card test

Memory card test

• Press <9>.

memory card test

• Insert memory card.

The memory card has to be formated from a 785 DMP Titrino otherwise the test does not recognize the card
and on the display appears „memory card 000 KB end“.

• Press .

The test runs automatically. If no fault arises then there appears:

memory card 128 KB o.k.

• Press .

diagnose press key 0...9

10 Dosing unit

Dosing unit

Dosing unit

Dosing unit

• Insert exchange unit and put the burette tip into a collecting receptacle.
• Press .

The Titrino gets out of the diagnosis menu and returns to the user program.

• Press , then .
• Remove exchange unit.
• Check spindle zero position, see Fig. 6.

0.2 - 0.6 mm


Fig. 6

Fig. 7

The spindle must be 0.2 - 0.6 mm below the edge of the sliding plate.

The bar of the stopcock coupling must be exactly parallel to the lateral edges of the
DMP Titrino, see Fig.7.

• Reinsert Exchange Unit.

Titrino

fills.

The display of before reappears.

• Knob 'dV/dt' to right stop.
• Press the key (on instrument) until the piston rod reaches the top and at the sa-

me time measure the time from start to end.

XXX X

********

cylinder empty !

mind selected language!