5.1 Setting up and connecting the instruments

785 DMP Titrino

153

5.1.4 Connection of a Sample Changer

With 6.2141.030 cable (instead of 6.2141.020), two Titrinos can be
connected to the 730 or 760 Sample Changer at the same time.

The "Remote" socket allows not only connection of a sample
changer but also additional control functions. Pin assignment of
the "Remote" socket and control possibilities, see page 164.

If a calibration has to be performed with the sample changer, the
calibration parameter "sample changer:" must be set to "ON".

In connections with the sample changer, "auto start" should be
set to "OFF" in the key. The start command is given
by the Sample Changer.

C

B

A

D

785

6.2141.020 Cable

730/760