BNC 1105 Universal Counter User Manual
Page 97
97
97
97
97
Channel 1 and 2
Trigger
Default setting is Auto Trigger at 50 %
“Auto” Gate Time
0.1 sec
STD CH 3
375 MHz to 6 GHz (0.166 ns to 2.6 ns)
Frequency Ratio
CH 1/ CH 2, CH 1/ CH 3, CH 2/ CH 1, CH 3/ CH1
(Measurement is specified over the full signal range of each input.)
Results Range
10E-10 to 10E+11
“Auto” Gate Time
0.1 sec
Time Interval
Measurement is specified over the full signal ranges of Channels 1 and 2. The width of the
pulse must be greater than 1 ns, frequency range to 300 MHz.
Trigger
Default setting is Auto Trigger at 50 %
Results Range
-0.5 ns to 10E+5 s
Resolution
40 ps
RMS Resolution
120 ps
Systematic
Uncertainty
±(TI × Time Base Error) ± Trigger Level Timing Error ± 500 ps Differential Channel Error
Pulse Width Time
Measurement is specified over the full signal range of Channel 1. The width of the pulse
must be greater than 1 ns frequency range to 300 MHz).
Pulse Selection
Positive or Negative
Trigger
Default setting is Auto Trigger at 50%
Results Range
1.5 ns to 10E+5 s
Resolution
40 ps
RMS Resolution
120 ps
Systematic
Uncertainty
± (Pulse Width Time x Time Base Error) ± Trigger Level Timing Error ± 500 ps Differential
Channel Error.
Rise/Fall Time
Measurement is specified over the full signal range of Channel 1. The width of the pulse
must be greater than 1 ns frequency range to 300 MHz).
Edge Selection
Positive or Negative
Trigger
Default setting is Auto Trigger at 10% and 90%
Results Range
2 ns to 10E+5 s
Resolution
40 ps
RMS Resolution
120 ps
Systematic
Uncertainty
± (Edge Time x Time Base Error) ± Trigger Level Timing Error ± 500 ps Differential
Channel Error
Phase
Measurement is specified over the full signal range of each input. The width of the pulse
must be greater than 1 ns, frequency range to 300 MHz
Results Range
-180° to +360°
Resolution
40 ps
RMS Resolution
120 ps
Systematic
Uncertainty
± (Trigger Level Timing Error) ×Frequency
Duty Cycle
Measurement is specified over the full signal range of Channel 1. The width of the pulse
must be greater than 1 ns, frequency range to 300 MHz
Pulse Selection
Positive or Negative
Trigger
Default setting is Auto Trigger at 50 %
Results Range
0 to 1
Resolution
40 ps
RMS Resolution
120 ps
Systematic
Uncertainty
± Trigger Level Timing Error ± 500 ps Differential Channel Error
Totalize
Measurement is specified over the full signal range of Channel 1. The width of the pulse
must be greater than 1 ns, frequency range to 400 MHz
Pulse Selection
Positive or Negative