:SYSTem:KEY:LOG?

Queries all the entries in the Key Queue.

Return

Numeric data transferred as ASCII bytes in format. A “-1” is returned when

the queue is empty. Each Key Code number is separated by a comma. The entries are

in a last-in, first-out order.

The maximum number of key codes is 500. Each key code has a range of 1 to 10.

Note

The Key Queue will be empties after the counter has been turned off or after a *RST

command.

Save/recall operation does not apply on this value.

:SYSTem:VERSion?

Queries the SCPI version number with which the counter complies.

Return

Numeric data transferred in ASCII bytes in format.

The response is in the form of YYYY.V, where YYYY represents the year and V

represents the approved version of that year.

:SYSTem:COMMunicate:RLSTate LOCal | REMote | RWLock

Set or queries the remote/local state of the instrument.

LOCal: Free the all panel keys.

REMote: Set I/O mode to remote and lock all of panel keys, but Preset/Local key not.

RWLock: To lock all of panel keys, instrument can only be controlled via the remote

interface).

SYSTem:BEEPer:STATe OFF | ON

Set or queries the beeper enable mode when an error is generated from the

Front panel or over the remote interface.

:SYSTem:IDNStr

Set the IDN String for I/O compatible mode. The setting string be queried by *idn? when