The command :calibration – BNC 1105 Universal Counter User Manual
Page 119

119
119
119
119
It will be returned to 0 after the counter has been turned off or a remote interface
reset.
When this is on, only the measurements (scaled and offset when math is on) within
limits will be used in statistical calculation.
:CALCulate3:PATH?
Queries the order in which CALCulate3 sub-blocks are to be processed.
Return
A sequence of ASCII-encoded bytes: LFIL, AVER.
Notes
For the counter, this sequence is fixed to be LFILter followed by AVERage.
The command :CALibration
:CALibration:SECurity:CODE
Sets the security code for calibration. The range is from 0 to 99,999,999, and
the resolution is 1. Default Security validate code is 1105.
Note
The security code is stored in non-volatile memory and is unaffected by
power-on, save/recall or *RST.
When the state is in the security code requirement, please input the correct
code. After that, if you want to change the code please input a new one. Then,
press the Enter button to store it.
:CALibration:SECurity:STATe
Sets or queries the security state of the calibration.
To unsecure the Counter for calibration, specify OFF with a valid security code.
Once the Counter is unsecured, any calibration can be performed.
To secure the Counter, specify ON with a valid security code. When the Counter
is secure against calibration, no calibration can be performed except for
interpolator calibration.
Return
Single ASCII-encoded byte. 0 means the Counter is unsecure while 1 means
the Counter is secure.