BNC 1105 Universal Counter User Manual

Page 138

background image

138

138

138

138

Return

The “100”, “50” or “0” PCT is returned.

Note

As the command *RST is executed, the noise immunity will be set to 50 PCT.

[:SENSe]EVENt[1|2]:LEVel[:ABSolute] [V]

Sets or queries the level of the center of the hysteresis window. The range is -5.125 to

5.125 volts for X1 attenuation and -51.25 to 51.25 volts for X10 attenuation. The

resolution is 0.0025V for X1 attenuation and 0.025V for X10 attenuation.

Return

Numeric data transferred as ASCII bytes in format with six significant digits.

Note

Sending this command disables [:SENS]:EVEN[1

︱2]:LEV:AUTO

[:SENSe]EVENt[1|2]:LEVel[:ABSolute]:AUTO

Sets or queries the state of the auto-triggering.

When auto-triggering is enabled, the counter automatically measures and computes the

trigger level according to the percentage specified by the command :[:SENS]:EVEN[1︱

2]:LEV:REL. Also the counter measures and computes trigger level each time :INIT

or :INIT:COUNT ON command is sent.

When the auto-triggering is on, the counter checks for each measurement if the

measurement signal is triggering, and the counter will measure and compute a new

trigger level if no triggering is found.

Return

A single ASCII-encoded byte. 0 means OFF and 1 means ON.

Notes

This will be set to ON after the counter has been turned off or a remote interface

reset.

Specifying a trigger level by using the command [:SENS]:EVEN[1

︱2]:LEV[:ABS] to

disable auto-triggering.

[:SENSe]EVENt[1|2]:LEVel:RELative [PCT]