[:SENSe]:FREQuency:ARM:STOP:DIGits
Sets or queries the resolution in terms of digits used in arming Period, Frequency and
frequency ratio measurements.
Return
Numeric data transferred as ASCII bytes in
resolution is 3 to 15.
Notes
This will be set to 4 after the counter has been turned off or a remote interface reset.
This command only applies when [:SENSe]:FREQ:ARM:STOP:SOUR DIG is used.
[:SENSe]:FREQuency:ARM:STOP:SLOPe POSitive
︱
︱
︱
︱NEGative
Sets or queries the slope of the external stop arm signal used in external arming for
Period, frequency and frequency ratio measurements.
Return
A sequence of ASCII-encoded bytes. POS means Positive and NEG means Negative.
Notes
This will be set to Negative after the counter has been turned off or a remote interface
reset.
This command only applies when [:SENSe]:FREQ:ARM:STOP:SOUR EXT is used.
[:SENSe]:FREQuency:ARM:STOP:SOURce IMMediate
︱
︱
︱
︱EXTernallll︱
︱
︱
︱TIMer︱
︱
︱
︱DIGits
Sets or queries the stop arm for Period, Frequency and Frequency ratio measurements.
Return
A sequence of ASCII-encoded bytes. IMM means Immediate, EXT means External,
TIM means Timer and DIG means Digits.
Note
This will be set to Timer after the counter has been turned off or a remote interface
reset.