interface reset.

[:SENSe]:FREQuency:ARM:STOP:DIGits

Sets or queries the resolution in terms of digits used in arming Period, Frequency and

frequency ratio measurements.

Return

Numeric data transferred as ASCII bytes in format, and the range of the

resolution is 3 to 15.

Notes

This will be set to 4 after the counter has been turned off or a remote interface reset.

This command only applies when [:SENSe]:FREQ:ARM:STOP:SOUR DIG is used.

[:SENSe]:FREQuency:ARM:STOP:SLOPe POSitive

︱NEGative

Sets or queries the slope of the external stop arm signal used in external arming for

Period, frequency and frequency ratio measurements.

Return

A sequence of ASCII-encoded bytes. POS means Positive and NEG means Negative.

Notes

This will be set to Negative after the counter has been turned off or a remote interface

reset.

This command only applies when [:SENSe]:FREQ:ARM:STOP:SOUR EXT is used.

[:SENSe]:FREQuency:ARM:STOP:SOURce IMMediate

︱EXTernallll︱

︱TIMer︱

︱DIGits

Sets or queries the stop arm for Period, Frequency and Frequency ratio measurements.

Return

A sequence of ASCII-encoded bytes. IMM means Immediate, EXT means External,

TIM means Timer and DIG means Digits.

Note

This will be set to Timer after the counter has been turned off or a remote interface

reset.