Sets or queries the start arm for Phase measurements.

Return

A sequence of ASCII-encoded bytes. IMM means Immediate and EXT means

External.

Note

This will be set to Immediate after the counter has been turned off or a remote

interface reset.

[:SENSe]:ROSCillator:EXTernal:FREQuency?

Under the timeBase measurement, (Fno) will be shown on the counter display if there is

no inputted frequency, or (Fxx) will be shown on the counter and “Not a number” on the

PC side if the inputted frequency is an invalid value. On the other hand, the command

ROSCillator:SOURce

will reply “Not a number” under the INTernal condition, and

“1.00000E+007” under the EXTernal condition.

Return

Numeric data transferred as ASCII bytes in format with six significant digits.

Note

If the current reference timebase is internal or if the current reference timebase is

external, but the frequency is not known, Not A Number 9.91E37 will be returned.

[:SENSe]:ROSCillator:SOURce INTernal|EXTernal

Sets or queries the current reference timebase.

Return

A sequence of ASCII-encoded bytes. INT means that the timebase is internal

reference. EXT means that external.

Notes

Specifying the reference timebase automatically disables

[:SENS]:ROSC:SOUR:AUTO ([:SENS]:ROSC:SOUR:AUTO OFF).

When [:SENS]:ROSC:SOUR:AUTO ON, this query tells which timebase reference is

used.

Save/recall operation does not apply on this value.