BNC 1105 Universal Counter User Manual
Page 149

149
149
149
149
negative edge.
Notes
This will be set to Positive after the counter has been turned off or a remote interface
reset.
This command only applies when [:SENSe]:TINT:ARM:ESTOP:LAY2:SOUR EXT is
used.
[:SENSe]:TINTerval:ARM:ESTOP:LAYer2:SLOPe IMMediate
︱
︱
︱
︱EXTernal
Sets or queries the stopping point of the arm for Time interval measurements.
Return
A sequence of ASCII-encoded bytes. IMM means Immediate and EXT means
External.
Note
This will be set to Immediate after the counter has been turned off or a remote
interface reset.
[:SENSe]:TINTerval:ARM:ESTOP[:LAYer[1]]:ECOunt
Sets or queries the number of events used to delay the stop arm for Time Interval
measurements. The range is from 1 to 99,999,999. The resolution is 1.
Return
Numeric data transferred as ASCII bytes in
Notes
This will be set back to 1 after the counter has been turned off or a remote interface
reset.
This command only applies when [:SENSe]:TINT:ARM:ESTOP[:LAYer[1]]:SOUR
Internal2 is used.
[:SENSe]:TINTerval:ARM:ESTOP[:LAYer[1]]:SOURce IMMediate
︱
︱
︱
︱ TIMer ︱
︱
︱
︱
INTernal2
Sets or queries the stopping point of the arm for Time Interval measurements.
Return