A single ASCII-encoded byte. 0 means that the result display is in enabled
and 0 means that the menu display is enabled.
Notes
The state of the display will be set back to OFF which means that the menu
display is disabled and the result display is enabled after *RST command.
Use :SYST:KEY command or the front panel key to display the menu.
This value is unaffected by save&recall.
:DISPlay[:WINDow]:TEXT:FEED “[:]CALCulate2”|”[:]CALCulate3”
Sets or queries what data flow is fed into the display.
“[:]CALCulate2”---is used to direct any result other than statistics to the result
display.
The result should be one of the following:
Raw measurement result when the mathematical function is not enabled
The scaled/offset measurement result when the mathematical function is
enabled
The limit graph when the limit testing is enabled and the graph mode is
chosen.
“[:]CALCulate3”---is used to direct the statistics result to the result display. The
particular statistic result displayed is determined by ::CALC3:AVER:TYPE.
Return
The string “CALC2” or “CALC3” is returned.
Note
This will be set back to “CALC2” after the *RST command.
:DISPlay[:WINDow]:TEXT:RADix COMMa|DPOint
Sets or queries the character used to separate integral and fractional portions of
a displayed number.
When using COMMa, one thousand is displayed as 1.000,0, but 1,000.0 if you
are using DPOint.
Return