Default
(@1)
Range
(@1)|(@2)
:MEASure[:SCALar][:VOLTage]:MINimum? [(@1)|(@2)]
Measures voltage minimum.
Default
(@1)
Range
(@1)|(@2)
:MEASure[:SCALar][:VOLTage]:NWIDth? [
Sets the measurement arming to “auto” and measures negative pulse width.
The
This point can be specified in percentage or in voltage. The default unit is percent.
The
in percentage or default, auto-trigger is enabled and trigger level is the specified value.
When it is specified in voltage, auto-trigger is disabled and the trigger level is the
specified value.
default
50 pct
Percent range
0 to 100 pct
Percent resolution
10%
Voltage range
-5.125V to 5.125V for ×1 attenuation
-51.25V to 51,25V for ×10 attenuation
Voltage resolution
0.0025V for ×1 attenuation
0.025V for ×10 attenuation
default
(@1)
range
(@1)
:MEASure[:SCALar][:VOLTage]:PWIDth? [
Sets the measurement arming to “auto” and measures positive pulse width.
The
This point can be specified in percentage or in voltage. The default unit is percent.
The
in percentage or default, auto-trigger is enabled and trigger level is the specified value.
When it is specified in voltage, auto-trigger is disabled and the trigger level is the
specified value.
default
50 pct
Percent range
0 to 100 pct
Percent resolution
10%