BNC 1105 Universal Counter User Manual

Page 133

background image

133

133

133

133

Default

(@1)

Range

(@1)|(@2)

:MEASure[:SCALar][:VOLTage]:MINimum? [(@1)|(@2)]

Measures voltage minimum.

Default

(@1)

Range

(@1)|(@2)

:MEASure[:SCALar][:VOLTage]:NWIDth? [][,(@1)]

Sets the measurement arming to “auto” and measures negative pulse width.

The specifies the point on the pulse where the negative pulse width is determined.

This point can be specified in percentage or in voltage. The default unit is percent.

The parameter is used to configure the trigger setting. When it is specified

in percentage or default, auto-trigger is enabled and trigger level is the specified value.

When it is specified in voltage, auto-trigger is disabled and the trigger level is the

specified value.

default

50 pct

Percent range

0 to 100 pct

Percent resolution

10%

Voltage range

-5.125V to 5.125V for ×1 attenuation

-51.25V to 51,25V for ×10 attenuation

Voltage resolution

0.0025V for ×1 attenuation
0.025V for ×10 attenuation

default

(@1)

range

(@1)

:MEASure[:SCALar][:VOLTage]:PWIDth? [][,(@1)]

Sets the measurement arming to “auto” and measures positive pulse width.

The specifies the point on the pulse where the positive pulse width is determined.

This point can be specified in percentage or in voltage. The default unit is percent.

The parameter is used to configure the trigger setting. When it is specified

in percentage or default, auto-trigger is enabled and trigger level is the specified value.

When it is specified in voltage, auto-trigger is disabled and the trigger level is the

specified value.

default

50 pct

Percent range

0 to 100 pct

Percent resolution

10%