BNC 1105 Universal Counter User Manual
Page 72

72
72
72
72
[:VOLTage]:PHASe? [(@1),(@2)]
[:VOLTage]:DCYCle? [
[:VOLTage]:PDUTycycle? [
[:VOLTage]:TOTalize:TIMed? [
The CONFigure Command
The CONFigure command offers a little more flexibility than the command
MEASure? and is usually used with the commands READ?. The universal
counter sets the parameters for the requested function, but does not make
measurement. Users have an option to change the configuration. To initiate
the measurement, READ? Or INITiate commands can be used.
:CONFigure[:SCALar]:
For more information about
refer to the above table.
•
If you skip the parameters or the function, or the source list, or use the
keyword DEFault instead, the default settings (parameters, function and
source) will be used.
•
Using this command automatically turns off the statistics operation, scale
and offset operation and limit testing.
•
If the trigger level is not specified in
be enabled and the trigger levels will be automatically set in all functions but
Voltage Peaks and Totalize on the measurement channel
.
:CONFigure
:CONFigure[:SCALar]
[:VOLTage]:FREQuency [
[,(@1)|(@2)|(@3)]
[:VOLTage]:FREQuency:RATio [
[,(@1),(@2)|(@3)]|[,(@2)|(@3),(@1)]
[:VOLTage]:PERiod [
[,(@1)|(@2)|(@3)]
[:VOLTage]:RISE:TIME [
[:VOLTage]:RTIMe [
[:VOLTage]:FALL:TIME [
[:VOLTage]:FTIMe [