The commands :trace – BNC 1105 Universal Counter User Manual
Page 156
156
156
156
156
I/O mode is compatible.
:SYSTem:UTILity:IO:MODE NATive
|
|
|
|COMPatibility
To set and query the IDN string, the NATive can be selected for default, and the
COMPatibility can be selected for user definition.
Return
A sequence of ASCII-encoded bytes: NAT or COMP.
Note
As the command *RST is executed, the NAT will be set.
The Commands :TRACe
These commands are used to specify and query the scale and offset
values.
:TRACe:CATalog?
Queries the names of the constants the counter uses.
Return
The strings “SCALE” and “OFFSET” are returned.
:TRACe[:DATA] OFFSET,
Or
:TRACe[:DATA] OFFSET,
Sets the offset value. The range for offset value is from -9.9999990000E+12 to
-1.0000000000E-13,
0.0000000000,
and
from
+1.0000000000E-13
to
+9.9999990000E+12. The resolution is 11 digits.
Notes
The offset value will be set back to zero after the counter has been turned off or after
a remote interface reset.
Specifying
a
new
offset
value
will
clear
the
limit
counts
(:CALC2:LIM:FCO, :CALC2:LIM:PCO).
:TRACe[:DATA]? OFFSET
Queries the offset value.