Numeric data transferred as ASCII bytes in format.

Notes

If the statistics operation is not enabled (:CALC3:AVER[:STATe] OFF, error -221 will

be generated.

If the current measurement operation is Totalize or Voltage Peaks, 0 will be returned

and error -221 will be generated.

:CALCulate3:AVERage[:STATe]

Sets or queries the statistics operation state.

Return

A single ASCII-encoded byte. 0 means OFF and 1 means ON.

Note

This will be set to OFF after the counter has been turned off or a remote interface

reset.

:CALCulate3:AVERage:TYPE MAXimum|MINimum|SDEViation|SCALar or MEAN

Sets the statistical result for :CALC3:DATA? Command and display

when :DISP[:WIND]:TEXT:FEED is set to “CALC3”.

Return

A sequence of ASCII-encoded bytes. There are MAX, MIN, SDEV, or MEAN.

Note

This will be set to its default, MEAN, after the counter has been turned off or a remote

interface reset.

:CALCulate3:DATA?

Queries the statistical result selection made by :CALC3:AVER:TYPE command.

Return

A result formatted by :FORMat[:DATA] ASCii|REAL command. If using the ASCII

format, it returns numeric data transferred in format. The number of