Numeric data transferred as ASCII bytes in
Notes
If the statistics operation is not enabled (:CALC3:AVER[:STATe] OFF, error -221 will
be generated.
If the current measurement operation is Totalize or Voltage Peaks, 0 will be returned
and error -221 will be generated.
:CALCulate3:AVERage[:STATe]
Sets or queries the statistics operation state.
Return
A single ASCII-encoded byte. 0 means OFF and 1 means ON.
Note
This will be set to OFF after the counter has been turned off or a remote interface
reset.
:CALCulate3:AVERage:TYPE MAXimum|MINimum|SDEViation|SCALar or MEAN
Sets the statistical result for :CALC3:DATA? Command and display
when :DISP[:WIND]:TEXT:FEED is set to “CALC3”.
Return
A sequence of ASCII-encoded bytes. There are MAX, MIN, SDEV, or MEAN.
Note
This will be set to its default, MEAN, after the counter has been turned off or a remote
interface reset.
:CALCulate3:DATA?
Queries the statistical result selection made by :CALC3:AVER:TYPE command.
Return
A result formatted by :FORMat[:DATA] ASCii|REAL command. If using the ASCII
format, it returns numeric data transferred in