:CALCulate3:LFILter:UPPer[:DATA]
Sets or queries the upper limit in statistics operation. The resolution is 11 digits and the
range is from -9.9999990000E+12 to -1.0000000000E-13, 0.0000000000, and
+1.0000000000E-13 to 9.9999990000E+12.
Return
Numeric data transferred in ASCII bytes in
digits.
Notes
It will return to its power-on default, 0, after the counter has been turned off or a
remote interface reset.
This command shares the same value with :CALC:LIM:UPP command.
When limit filtering is enabled (:CALC:LFIL:STAT ON), any value higher than this
value will not be used in the statistics calculation.
If math operation is enabled (:CALC3:MATH:STATe ON), user should consider the
scaled and offset measurement when specifying the limit.
Setting
a
new
upper
limit
will
clear
the
limit
counts
(:CALC2:LIM:FCO, :CALC2:LIM:PCO).
When the display is not able to show all of the significant digits of this value, the
displayed value is different from the actual value because the displayed value has
been rounded. But if you press ENTER while the number is still in display, the actual
value will be updated to the value shown.
:CALCulate3:LFILter:STATe
Sets or queries the state of the statistics filtering.
Return
A single ASCII-encoded byte. 1 means ON and 0 means OFF.
Notes