receives :INIT[:IMM].

When :CALC2:LIM:CLE:AUTO is ON and whenever the counter

receives :INIT:CONT ON command, it :

o

Invalidates the limit data,

o

Clears the pass and fail counts (:CLAC2:LIM:FCO and :CALC2:LIM:PCO)

in the limit testing.

o

Sets the Limit-Detect output to the in-limit voltage level.

When [:SENSe]:EVEN:LEV[:ABS]:AUTO is ON and :INIT[:IMM] command is

sent, the counter issues an auto-trigger on the source.

If the counter is in the process of making a measurement or

if :iNItiate:CONTinuous is set to ON, an :IMMediate command has no effect,

and error -231 (Init ignored) is generated.

Beginning a measurement or block of measurements with an :INIT[:IMM]

sets the pending Operation Flag to true. Completing the measurement or

block of measurements (normally or by aborting) sets Pending Operation

Flag to false.

For an automatically armed Totalize measurement, use :ABORt to terminate

the measurement initiated with :INIT[:IMM].

:INPut[1|2]:ATTenuation 1|10

Sets or queries the input attenuation.

Return

Numeric data transferred as ASCII bytes in format.

Note

The input attenuation is set to X1 after *RST command.

:INPut[1|2]:COUPling AC|DC

Sets or queries the input coupling.

Return

A sequence of ASCII-encoded bytes: AC or DC.

Note

The input coupling is set to AC after *RST command.

:INPut[1|2]:FILTer[:LPASs][:STATe]