beautypg.com

BNC 1105 Universal Counter User Manual

Page 132

background image

132

132

132

132

Resolution should use a mantissa of 1.0 and be an even

power of 10

Ch3

Default

Value which indicates 4 digits of resolution for the specified

Range

1E-7 Hz ~ 1E7 Hz (15 to 3 digits of resolution)

Resolution should use a mantissa of 1.0 and be an even

power of 10

Default

(@1)

Range

(@1|@2|@3)

:MEASure[:SCALar][:VOLTage]:FREQuency:RATio?

[ [,]][,(@1),(@2)|(@1),(@3)|(@2),(@1)|(@3),(@1)]

Sets the measurement arming to “digits” and measures the frequency ratio between two inputs.

The and parameters are used to configure the number of digits

of resolution arming setting.

default

1

Range

Ch1/Ch2, Ch1/Ch3

1E-10 to 1E11

Range

Ch2/Ch1, Ch3/Ch1

1E-11 to 1E10

Resolution

should be within

10% of ratio for optimum arming

configuration

Default

Value which indicates 4 digits of

resolution for the specified

value>

Range

Ch1/Ch2, Ch1/Ch3

1E-25 ~ 1E8
(15 to 3 digits of resolution)

Range

Ch2/Ch1, Ch3/Ch1

1E-26 ~ 1E7
(15 to 3 digits of resolution)

Resolution

should use a mantissa of

1.0 and be an even power of 10

Default

(@1,@2)

Range

(@1,@2)|(@1,@3)|(@2,@1)|(@3,@1)

:MEASure[:SCALar][:VOLTage]:MAXimum? [(@1)|(@2)]

Measures voltage maximum.