A sequence of ASCII-encoded bytes. IMM means Immediate, INT means Internal2

and TIM means Timer.

Note

This will be set to Immediate after the counter has been turned off or a remote

interface reset.

This command does not apply on Rise/Fall time, Pulse width or Duty cycle

measurements.

[:SENSe]:TINTerval:ARM:ESTop[:Layer[1]]:TIMer [S]

Sets or queries the delay time used in stop arm for Time Interval measurements. The

range is 100E-9 to 10.000 seconds and the resolution is 100E-9 seconds.

Return

Numeric data transferred as ASCII bytes in format.

Notes

This will be set to 100E-9 seconds after the counter has been turned off or a remote

interface reset.

This command only applies when [:SENSe]:TINT:ARM:ESTOP[:LAY[1]]:SOUR TIM is

used.

This command does not apply on Rise/Fall Time, Pulse width or Duty cycle

measurements.

Sets or queries

Gate

for Totalize measurements. For more information about

the front

panel gating settings and the interface command parameters, please refer to

the TOT table in Chapter 5.4.

[:SENSe]:TOTalize:ARM[:START]:SLOPe POSitive

︱NEGative

Sets or queries the slope of the external start arm signal used in external arming Totalize

measurements.

Return

A sequence of ASCII-encoded bytes. POS means Positive and NEG means Negative.

Notes

This will be set to Positive after the counter has been turned off or a remote interface

reset.