significant digits will range from 1 to 15, depending on the resolution.

Notes

Be sure to enable the statistics operation before query the result.

If the statistics operation is not enabled (:CALC3:AVER[:STATe] OFF), or the current

measurement function is Totalize or Voltage Peaks, Not A Number 9.91E37 will be

returned and error -221 will be generated.

If there is no valid result, Not A Number 9.91E37 will be returned and error -230 will

be generated.

The last calculated statistical result remains valid until a new calculation is made or a

relevant instrument state is modified.

:CALCulate3:FEED “[:]CALCulate[1]”

Sets or queries the data flow to be fed into the CALCulate3 block.

Return

The string “CALC” is returned.

:CALCulate3:LFILter:LOWer[:DATA] [HZ|S|DEG]

Sets or queries the lower limit in statistics operation. The resolution is 11 digits and the

range is from -9.9999990000E+12 to -1.0000000000E-13, 0.0000000000, and

+1.0000000000E-13 to 9.9999990000E+12.

Return

Numeric data transferred in ASCII bytes in format with eleven significant

digits.

Notes

It will return to its power-on default, 0, after the counter has been turned off or a

remote interface reset.

This command shares the same value with :CALC:LIM:LOW command.

When limit filtering is enabled( :CALC:LFIL:STAT ON), any value lower than this

value will not be used in the statistics calculation.

If math operation is enabled(:CALC3:MATH:STATe ON), user should consider the

scaled and offset measurement when specifying the limit.

Setting

a

new

lower

limit

will

clear

the

limit

counts