BNC 1105 Universal Counter User Manual
Page 121

121
121
121
121
A string with the form of “
returned.
If there is not any specification before this query, an empty string is returned
and an error -221,"Settings conflict" is generated.
The response is unaffected by *RST, recall, or [:SENS]:FUNC.
Note
Refer to (The Command :MEASure?) later in this section for the descriptions
of the
Device Clear
Instructs the counter to clear the input buffer and Output Queue, resets
the parser, execution control and response formatter. Clears any
command that would prevent processing a command *RST or others.
Disables the effect of previously executed the command *OPC and
terminates the hold-off action of a *WAI, *OPC? Or data query
(:MEASure query, :READ query, :FETCh
query, :CALC:DATA?, :CALC2:LIM:FAIL?) waiting for pending operation
to complete.
Note: Diagnostics or calibration initiated from front panel may be terminated it
is waiting for user input.
:DISPlay:ENABle
Sets or queries if the whole display is enabled.
Return
A single ASCII-encoded byte. 0 means Disables and 1 means Enabled.
Notes
The display will be set back to Enabled after *RST command.
This value is unaffected by save&revall.
:DISPlay:MENU[:STATe] OFF
Disables the menu display and shows the results. Response indicates whether
menu or results is in display.