BNC 1105 Universal Counter User Manual
Page 146

146
146
146
146
[:SENSe]:ROSCillator:SOURce:AUTO
Sets or queries the state of the auto-specifying reference timebase. With this on, the
counter automatically choose the external reference signal as the reference timebase
when a valid signal is present at the Ref In connector on the rear panel. When an invalid
signal is present at this connector, the internal timebase is used.
Return
A single ASCII-encoded byte. 0 means Off and 1 means ON.
Notes
Specifying a reference oscillator with [:SENS]:ROSC:SOUR INT
︱EXT disables this
auto-specifying.
This will be set back to its default, which is ON, after the counter has been turned off
or after a remote interface reset.
Sets or queries
Gate
for Risetime, Falltime, Duty Cycle, and Pulse Width functions
measurements. For more information about
the front panel gating settings and the
interface command parameters, please refer to the Rise/Fall form in Chapter
5.4.
[:SENSe]:TINTerval:ARM:ESTART:LAYer2:SOURce IMMediate | EXTernal
Sets or queries the start arm for Risetime, Falltime, Duty Cycle, and Pulse Width
functions measurements.
Query Response:
A sequence of ASCII-encoded bytes: IMM or EXT
Comments:
*RST: IMMediate
Related:
Gate & ExtArm
[:SENSe]:TINTerval:ARM:ESTART:LAYer2:SLOPe POSitive | NEGative
Sets or queries the slope of the external start arm signal used in external arming
Risetime, Falltime, Duty Cycle, and Pulse measurements.
Query Response:
A sequence of ASCII-encoded bytes: POS or NEG
Comments:
● *RST: POSitive
● Only applies when [:SENS]:TINT:ARM:ESTART:LAY2:SOUR EXT is
selected.
Related:
Gate & ExtArm
Sets or queries
Gate
for Time interval measurements. For more information about
the