The command :configure – BNC 1105 Universal Counter User Manual
Page 120

120
120
120
120
Notes
The security state is stored in non-volatile memory and is unaffected by
power-on, save/recall or *RST.
The security code is set to 1105 when the Counter is shipped from the
factory. If users forget the security code, they can use
the :CALibration:SECurity:CODE:RESet command to recover to the default
condition.
:CALibration:SECurity:CODE:RESet
To prevent missing the user-defined code, and access the calibration mode
quickly, users can execute the command to recover to the default condition.
The Command :CONFigure
The commands :CONFigure and :READ offer more flexibility than the
command :MEASure. It provides the configuration part of the
measurement, but does not perform measurement nor query the result.
When using the comand :CONFigure, the Scale and Offset operation,
statistics and limit-testing are all disabled. If an absolute trigger level is
not specified in the
will be enabled and set for all measurement function except Totalize and
Voltage Peaks. Refer to (The Command :MEASure?) later in this section
for the descriptions of the
function.
※Note: When you use this command, several default settings are automatically
used. In order to switch to another function without changing the settings, use
[:SENS]:FUNC[:ON]
:CONFigure
[:SCALar]:
To perform the specified function without initiating the measurement, please use the
command :INIT;FETC? or READ? to make and query a measurement. Skipping the
:CONFigure?
Queries the function configured by the last :CONFigure or :MEASure command.
Return