BNC 1105 Universal Counter User Manual
Page 151
151
151
151
151
This command only applies when [:SENSe]:TOT:ARM[:STAR]:SOUR EXT is used.
[:SENSe]:TOTalize:ARM[:STARt]:SOURce IMMediate
︱
︱
︱
︱EXTernallll
Sets or queries the starting point of the arm for Totalize measurement.
Return
A sequence of ASCII-encoded bytes. IMM means Immediate and EXT means
External.
Notes
This will be set to Immediate after the counter has been turned off or a remote
interface reset.
When the staring and the stopping points are both set to immediate, the :INIT[:IMM]
command can be used to start totalizing and the :ABORt command can be used to
end it. Before the “auto” totalize measurement is terminated, there is no valid
totalize result.
[:SENSe]:TOTalize:ARM:STOP:SLOPe POSitive
︱
︱
︱
︱NEGative
Sets or queries the slope of the external stop arm signal used in external arming Totalize
measurements.
Return
A sequence of ASCII-encoded bytes. POS means Positive and NEG means Negative.
Notes
This will be set to Negative after the counter has been turned off or a remote interface
reset.
This command only applies when [:SENSe]:TOT:ARM:STOP:SOUR EXT is used.
[:SENSe]:TOTalize:ARM:STOP:SOURce IMMediate
︱
︱
︱
︱TIMer︱
︱
︱
︱EXTernallll
Sets or queries the stopping point of the arm for Totalize measurements.
Return
A sequence of ASCII-encoded bytes. IMM means Immediate, TIM means Timer and
EXT means External.
Notes
This will be set to Timer after the counter has been turned off or a remote interface