This command only applies when [:SENSe]:TOT:ARM[:STAR]:SOUR EXT is used.

[:SENSe]:TOTalize:ARM[:STARt]:SOURce IMMediate

︱EXTernallll

Sets or queries the starting point of the arm for Totalize measurement.

Return

A sequence of ASCII-encoded bytes. IMM means Immediate and EXT means

External.

Notes

This will be set to Immediate after the counter has been turned off or a remote

interface reset.

When the staring and the stopping points are both set to immediate, the :INIT[:IMM]

command can be used to start totalizing and the :ABORt command can be used to

end it. Before the “auto” totalize measurement is terminated, there is no valid

totalize result.

[:SENSe]:TOTalize:ARM:STOP:SLOPe POSitive

︱NEGative

Sets or queries the slope of the external stop arm signal used in external arming Totalize

measurements.

Return

A sequence of ASCII-encoded bytes. POS means Positive and NEG means Negative.

Notes

This will be set to Negative after the counter has been turned off or a remote interface

reset.

This command only applies when [:SENSe]:TOT:ARM:STOP:SOUR EXT is used.

[:SENSe]:TOTalize:ARM:STOP:SOURce IMMediate

︱TIMer︱

︱EXTernallll

Sets or queries the stopping point of the arm for Totalize measurements.

Return

A sequence of ASCII-encoded bytes. IMM means Immediate, TIM means Timer and

EXT means External.

Notes

This will be set to Timer after the counter has been turned off or a remote interface