:MEASure[:SCALar][:VOLTage]:FALL:TIME?

[ [,upper_reference>]][,(@1)]

Sets the measurement arming to “auto” and measures Fall Time with the specified

parameters.

The and are used to specify the lower and upper

points on the falling edge of the input signal. They can be specified as either a percentage

or an absolute voltage.

The and are used to configure the trigger setting.

When they are in percentage or defaults, auto-trigger is enabled and trigger level is the

specified value. When they are specified in voltage, auto-trigger is disabled and the

trigger level is the specified value.

default

10 pct

Percent range

0 to 100 pct

Percent resolution

10%

Voltage range

-5.125V to 5.125V for ×1 attenuation

-51.25V to 51,25V for ×10 attenuation

Voltage resolution

0.0025V for ×1 attenuation
0.025V for ×10 attenuation

default

90 pct

Percent range

0 to 100 pct

Percent resolution

10%

Voltage range

-5.125V to 5.125V for ×1 attenuation

-51.25V to 51,25V for ×10 attenuation

Voltage resolution

0.0025V for ×1 attenuation
0.025V for ×10 attenuation

default

(@1)

range

(@1)

:MEASure[:SCALar][:VOLTage]:RISE:TIME?

[ [,upper_reference>]][,(@1)]

Sets the measurement arming to “auto” and measures Rise Time with the specified

parameters.

The and are used to specify the lower and upper

points on the falling edge of the input signal. They can be specified as either a percentage

or an absolute voltage.

The and are used to configure the trigger setting.

When they are in percentage or defaults, auto-trigger is enabled and trigger level is the