BNC 1105 Universal Counter User Manual
Page 130
130
130
130
130
:MEASure[:SCALar][:VOLTage]:FALL:TIME?
[
Sets the measurement arming to “auto” and measures Fall Time with the specified
parameters.
The
points on the falling edge of the input signal. They can be specified as either a percentage
or an absolute voltage.
The
When they are in percentage or defaults, auto-trigger is enabled and trigger level is the
specified value. When they are specified in voltage, auto-trigger is disabled and the
trigger level is the specified value.
default
10 pct
Percent range
0 to 100 pct
Percent resolution
10%
Voltage range
-5.125V to 5.125V for ×1 attenuation
-51.25V to 51,25V for ×10 attenuation
Voltage resolution
0.0025V for ×1 attenuation
0.025V for ×10 attenuation
default
90 pct
Percent range
0 to 100 pct
Percent resolution
10%
Voltage range
-5.125V to 5.125V for ×1 attenuation
-51.25V to 51,25V for ×10 attenuation
Voltage resolution
0.0025V for ×1 attenuation
0.025V for ×10 attenuation
default
(@1)
range
(@1)
:MEASure[:SCALar][:VOLTage]:RISE:TIME?
[
Sets the measurement arming to “auto” and measures Rise Time with the specified
parameters.
The
points on the falling edge of the input signal. They can be specified as either a percentage
or an absolute voltage.
The
When they are in percentage or defaults, auto-trigger is enabled and trigger level is the