Electrical connections – KROHNE IFC 300 Converter EN User Manual

Page 91

background image

ELECTRICAL CONNECTIONS

4

91

IFC 300

www.krohne.com

08/2010 - 4000069803 - MA IFC 300 R04 en

Status output / limit switch active, modular I/Os

• Observe connection polarity.
• U

int

= 24 VDC

• I ≤ 20 mA
• R

L

≤ 47 kΩ

• open:

I ≤ 0.05 mA
closed:
U

0, nom

= 24 V at I = 20 mA

• X designates the connection terminals A, B or D, depending on the version of the signal

converter.

Status output / limit switch passive, modular I/Os

• Any connection polarity.
• U

ext

= 32 VDC

• I ≤ 100 mA
• R

L, max

= 47 kΩ

R

L, min

= (U

ext

- U

0

) / I

max

• open:

I ≤ 0.05 mA at U

ext

= 32 VDC

closed:
U

0, max

= 0.2 V at I ≤ 10 mA

U

0, max

= 2 V at I ≤ 100 mA

• The output is open when the device is de-energized.
• X designates the connection terminals A, B or D, depending on the version of the signal

converter.

Figure 4-54: Status output / limit switch active S

a

Figure 4-55: Status output / limit switch passive S

p