Operation – KROHNE IFC 300 Converter EN User Manual
Page 131

OPERATION
6
131
IFC 300
www.krohne.com
08/2010 - 4000069803 - MA IFC 300 R04 en
C5. .11
format 3.line
Specify decimal places.
Select: automatic (adaptation carried out automatically) /
X (= none) …X.XXXXXXXX (max. 8 digits)
C5.5 graphic page
C5.5
graphic page
Graphic page always shows trend curve of the measurement of the 1. meas.
page / 1.line, see Fct. C5.3.2
C5.5.1
select range
Select: manual (set range in Fct. C5.5.2) /
automatic (automatic depiction based on the measured values)
Reset only after parameter change or after switching off and on.
C5.5.2
range
Set the scaling for the Y axis. Only available if "manual" is set in C5.5.1.
+xxx ±xxx%; range: -100…+100%
(1st value = lower limit / 2nd value = upper limit),
condition: 1st value ≤ 2nd value
C5.5.3
time scale
Set the time scaling for the X axis, trend curve
xxx min; range: 0…100 min
C5.6 special functions
C5.6
special functions
-
C5.6.1
reset errors
reset errors?
Select: no/yes
C5.6.2
save settings
Save current settings.
Select: break (exit function without saving) / backup 1 (save in storage
location 1) / backup 2 (save in storage location 2)
Query: continue copy? (cannot be done afterwards)
Select: no (exit function without saving) / yes (copy current settings to
storage backup 1 or backup 2)
C5.6.3
load settings
Load saved settings.
Select: break (exit function without loading) / factory settings (load in state
as delivered) / backup 1 (load data from storage location 1) / backup 2 (load
data from storage location 2) / load sensor data (factory settings of
calibration data)
Query: continue copy? (cannot be done afterwards)
Select: no (exit the function without saving) / yes (load data from the selected
storage location)
C5.6.4
password quick set
Password required to change data in the quick setup menu.
0000 (= to Quick Setup menu without password)
xxxx (password required); range 4digit: 0001…9999
C5.6.5
password setup
Password required to change data in the setup menu.
0000 (= to Quick Setup menu without password)
xxxx (password required); range 4digit: 0001…9999
C5.6.6
GDC IR interface
After this function has been activated an optical GDC adapter can be
connected to the LC display. If approximately 60 seconds pass without a
connection being established or after the adapter is removed, then the
function is exited and the optical keys are active once again.
Select:
break (exit function without connection) /
activate (the IR interface (adapter) and interrupt the optical keys)