OPERATION

6

131

IFC 300

www.krohne.com

08/2010 - 4000069803 - MA IFC 300 R04 en

C5.….11

format 3.line

Specify decimal places.

Select: automatic (adaptation carried out automatically) /

X (= none) …X.XXXXXXXX (max. 8 digits)

C5.5 graphic page

C5.5

graphic page

Graphic page always shows trend curve of the measurement of the 1. meas.

page / 1.line, see Fct. C5.3.2

C5.5.1

select range

Select: manual (set range in Fct. C5.5.2) /

automatic (automatic depiction based on the measured values)

Reset only after parameter change or after switching off and on.

C5.5.2

range

Set the scaling for the Y axis. Only available if "manual" is set in C5.5.1.

+xxx ±xxx%; range: -100…+100%

(1st value = lower limit / 2nd value = upper limit),

condition: 1st value ≤ 2nd value

C5.5.3

time scale

Set the time scaling for the X axis, trend curve

xxx min; range: 0…100 min

C5.6 special functions

C5.6

special functions

-

C5.6.1

reset errors

reset errors?

Select: no/yes

C5.6.2

save settings

Save current settings.

Select: break (exit function without saving) / backup 1 (save in storage

location 1) / backup 2 (save in storage location 2)

Query: continue copy? (cannot be done afterwards)

Select: no (exit function without saving) / yes (copy current settings to

storage backup 1 or backup 2)

C5.6.3

load settings

Load saved settings.

Select: break (exit function without loading) / factory settings (load in state

as delivered) / backup 1 (load data from storage location 1) / backup 2 (load

data from storage location 2) / load sensor data (factory settings of

calibration data)

Query: continue copy? (cannot be done afterwards)

Select: no (exit the function without saving) / yes (load data from the selected

storage location)

C5.6.4

password quick set

Password required to change data in the quick setup menu.

0000 (= to Quick Setup menu without password)

xxxx (password required); range 4digit: 0001…9999

C5.6.5

password setup

Password required to change data in the setup menu.

0000 (= to Quick Setup menu without password)

xxxx (password required); range 4digit: 0001…9999

C5.6.6

GDC IR interface

After this function has been activated an optical GDC adapter can be

connected to the LC display. If approximately 60 seconds pass without a

connection being established or after the adapter is removed, then the

function is exited and the optical keys are active once again.

Select:

break (exit function without connection) /

activate (the IR interface (adapter) and interrupt the optical keys)