8.5 Measuring accuracy (except TIDALFLUX)

Reference conditions

• Medium: water
• Temperature: 20°C / 68°F
• Pressure: 1 bar / 14.5 psi
• Inlet section: ≥ 5 DN

X [m/s]: flow velocity
Y [%]: deviation from the actual measured value (mv)

1

2

3

DN [mm]

DN [inch]

Accuracy

Curve

OPTIFLUX 5300

10….100

3/8…4

0.15% of mv + 1 mm/s

1

150...300

6...12

0.2% of mv + 1 mm/s

2

OPTIFLUX 2300 / 4300 / 6300

10….1600

3/8…80

0.2% of mv + 1 mm/s

2

OPTIFLUX 1300

10…150

3/8…6

0.3% of mv + 2 mm/s

3

OPTIFLUX 2300 / 4300

>1600

>64

0.3% of mv + 2 mm/s

3

OPTIFLUX 4300 / 5300 / 6300

<10

<3/8

0.3% of mv + 2 mm/s

3

OPTIFLUX 7300

25...100

1...4

v ≥ 1 m/s / 3.3 ft/s:
±

0.5% of mv

-

v < 1 m/s / 3.3 ft/s:
±

0.5% of mv + 5 mm/s

WATERFLUX 3300

25...600

1...24

0.2% of mv + 1 mm/s

2