Operation, 5 status messages and diagnostic information – KROHNE IFC 300 Converter EN User Manual

Page 135

background image

OPERATION

6

135

IFC 300

www.krohne.com

08/2010 - 4000069803 - MA IFC 300 R04 en

6.5 Status messages and diagnostic information

Operational faults in device

Messages on the display Description

Actions

Status: F _ _ _ _ _

Operational fault in device, mA output ≤
3.6 mA or set fault current (depending
on the seriousness of the fault), status
output open, pulse / frequency output:
no pulses

Repair necessary.

F error in device

Fault or failure of device. Parameter or

hardware error. No measurement possible.

Group message, when one of the following

or some other severe error occurs.

F IO 1

Error, operational fault in IO 1. Parameter

or hardware error. No measurement

possible.

Load settings (Fct. C4.6.3) (backup 1,

backup 2 or factory settings). If status

message still does not disappear, replace

electronic unit.

F parameter

Error, operational fault of data manager,

electronic unit, parameter or hardware

error. Parameters no longer usable.

F IO 2

Error, operational fault in IO 2. Parameter

or hardware error. No measurement

possible.

F configuration (also when

changing modules)

Invalid configuration: display software, bus

parameter or main software do not match

existing configuration. This error also

occurs when a module has been added or

removed without confirming the

configuration change.

After module change, confirm query for

changed configuration. If device

configuration unchanged: defective, replace

electronic unit.

F display

Error, operational fault in display.

Parameter or hardware error. No

measurement possible.

Defective, replace electronic unit.

F sensor electronic

Error, operational fault in sensor

electronics. Parameter or hardware error.

No measurement possible.

Defective, replace electronic unit.

F sensor global

Data error in the global data of the

measuring sensor electronic equipment.

Load settings (Fct. C5.6.3) (backup 1,

backup 2 or factory settings). If status

message still does not disappear, replace

electronic unit.

F sensor local

Data error in the local data of the measuring

sensor electronic equipment.

Defective, replace electronic unit.

F field current local

Data error in the local data of the field

current supply

Defective, replace electronic unit.

F current in-/output A

Error, operational fault in current output or

output for terminals A/B. Parameter or

hardware error. No measurement possible.

Defective, replace electronic unit or

input/output module (I/O module).

F current in-/output B

F current output C

Error, operational fault in current output for

terminal C. Parameter or hardware error.

Measurement not possible.

Defective, replace electronic unit or output

module (I/O module).

F software user interface

Fault revealed by CRC check of operation

software.

Replace electronic unit.

F hardware settings (also

when changing modules)

The set hardware parameters do not match

the identified hardware. A dialogue appears

in the display.

Answer queries in dialogue mode, follow

directions.

After module change, confirm query for

changed configuration. If device

configuration unchanged: defective, replace

electronic unit.

F hardware detection

Existing hardware cannot be identified.

Defective or unknown modules.

Replace electronic unit.