Operation – KROHNE OPTIWAVE 5200 C_F EN User Manual

Page 87

background image

OPERATION

6

87

OPTIWAVE 5200 C/F

www.krohne.com

07/2013 - 4001904902 - HB OPTIWAVE 5200 R02 en

ERR 103 Converter memory

failure

1

The device's hardware is

defective.

Replace the signal converter. For more

data, refer to

How to turn or remove the

signal converter

on page 37.

ERR 104 Converter Voltage

failure

1

The device's hardware is

defective.

Replace the signal converter. For more

data, refer to

How to turn or remove the

signal converter

on page 37.

ERR 200 Internal

Communication

1

The device's hardware or

software is defective. The

converter cannot transmit

signals to or receive signals

from the antenna

electronics.

De-energize the device. Make sure that the

signal cable engages in the terminal and

the screw connection is tight. Energize the

device. If the problem continues, replace

the signal converter. For more data, refer

to

How to turn or remove the signal

converter

on page 37.

ERR 201 Overfill

2

The level is in the blocking

distance. There is a risk that

the product will overflow

and/or cover the antenna.

Use a different procedure to measure the

level in the tank. Remove some of the

product until the level is below the blocking

distance.

If a viscous product touched the antenna:

remove the device and clean the
antenna, or

If the device has a purging option, use
the purging system to clean the
antenna.

ERR 203 Peak Lost (Level Lost)

2

The signal peak is not found

within the measuring

window that filters the

signals received by the

antenna. The measurement

is not correct. The device

will automatically increase

this window to find the

correct signal.

Do a check of the device, tank and the

process. Reconfigure the device and record

a new empty spectrum. Follow the

instructions on page 82. If necessary,

speak to the supplier.

ERR 204 Sensor no Signal

3

The device's hardware is

defective.

Replace the signal converter.For more

data, refer to

How to turn or remove the

signal converter

on page 37.

ERR 206 Sensor Microwave

failure

1

The device's hardware is

defective.

Replace the signal converter.For more

data, refer to

How to turn or remove the

signal converter

on page 37.

ERR 207 Sensor Memory

failure

1

The device's hardware is

defective.

Replace the signal converter.For more

data, refer to

How to turn or remove the

signal converter

on page 37.

ERR 208 Sensor Voltage failure

3

The device's hardware is

defective.

Do a check of the power supply at the

device terminals. Make sure that voltage

values are in the specified limits in menu

item 2.2.2 DIAGNOSTIC (Configuration

mode / Supervisor menu). If the voltage is

correct, replace the signal converter. For

more data on how to replace the signal

converter, refer to

How to turn or remove

the signal converter

on page 37.

ERR 210 Sensor Not

compatible

1

The software version of the

sensor is not compatible

with the software version of

the signal converter.

Go to menu 1.1.0 IDENT. in Configuration

mode. Record the version numbers of the

device software given in menu items 1.1.2,

1.1.3 and 1.1.4. Give this data to the

supplier.

1

Defective wiring.

Error

code

Error Message

Status

marker

shown

Cause

Corrective action