How to change from multidrop to point-to-point mode
• Enter the supervisor menu.
• Push [>>>>], 5 × [
] and [>>>>] to go to menu item ADDRESS (2.6.1).
• Push [>>>>] to change the value. Enter the value 0 and push [^
^
^
^] to confirm.
• Push [^
^
^
^] to go back to the "STORE" screen.
• Push [
] or [
] to set the screen to STORE YES
STORE YES
STORE YES
STORE YES and push [^
^
^
^].
i
The output is set to point-to-point mode. The current output changes to a range of 4...20 mA
or 3.8...20.5 mA (this range is set in menu item RANGE I (2.4.2)).
6.4.5 Distance measurement
The device displays distance measurements when the output function is set to
"Distance". Menu items related to distance measurement are:
• Output Function (2.4.1)
• Tank Height (2.3.1)
• Blocking Distance (2.3.2)
For more data about the menu items, refer to
Function description
on page 68, table 2.
Supervisor menu.
CAUTION!
If the distance for 4 mA is set in the blocking distance, the device will not be able to use the full
current output range.
Figure 6-4: Distance measurement
1 Tank Height (2.3.1)
2 Blocking Distance (2.3.2)
3 4 mA Setting (2.4.3)
4 20 mA Setting (2.4.4)
5 Maximum effective measuring range
6 Non-measurement zone