OPERATION

6

73

OPTIWAVE 5200 C/F

www.krohne.com

07/2013 - 4001904902 - HB OPTIWAVE 5200 R02 en

2.5.7

EMP.SPEC.EN.

The supervisor can start or stop the empty

spectrum filter.

YES, NO

NO

2.6.0 COMMUNICAT.

2.6.1

ADDRESS

Any HART

®

address greater than 0 will

activate HART

®

multidrop mode. The current

output stays constant at 4 mA.

min.-max: 0…63

0

2.7.0 DISPLAY

2.7.1

LANGUAGE

Data can be shown in any of the languages

stored in the device.

9 languages are

available in 3 packs:

(1) English, French,

German and Italian;

(2) English, French,

Spanish and

Portuguese;

(3) English, Chinese

(Mandarin), Japanese

and Russian

3

2.7.2

LENGTH UNIT

The length unit shown in normal mode.

m, cm, mm, in (inches),

ft (feet)

m

2.7.3

CONV UNIT

Conversion unit. The length, volume, mass or

flow rate conversion unit for the conversion

table and shown in normal mode.

m3, L, gal (US gallons),

ImpG (Imperial

gallons), ft3, bbl (oil

barrel), kg, t, Ston,

Lton, m, cm, mm, in, ft,

m3/h, ft3/h

L

2.7.4

PSWD YES/NO

If it is necessary to protect your settings in

the supervisor menu with a password, set

this menu item to YES

YES

YES

YES.

YES, NO

YES

2.7.5

PASSWORD

This changes the password for the supervisor

menu. Push the buttons up to 6 times in any

order. This will be the new password. To

confirm the change, enter the new password

a second time. For more data, refer to

Protection of the device settings

on page 78.

[>>>>], [^

^

^

^], [

],

[

], [>>>>] and

[^

^

^

^]

2.7.6

CONTRAST

The contrast control for the display screen.

You can select a shade of grey between light

grey (level 20) and black (level 54).

min.-max: 20…54

36

2.8.0 CONV. TABLE

2.8.1

INPUT TABLE

The device uses a conversion table (strapping

table) to convert measurements to volume,

mass and flow rate readings. The readings

are shown in normal mode. Give the number

of entries on the table (min. 2; max. 30).

Select an entry (01...30) and enter the level

and the related volume / mass / flow rate

value for that entry. Push [^

^

^

^] to confirm the

entry values. Continue the procedure until

the device has data for all the entries. For

more data, refer to

How to configure the

device to measure volume or mass

on page

81.

min. 2 entries

max. 30 entries

(level / volume, mass

or flow rate)

0 entries

2.8.2

DELETE TABLE

This menu item erases the data in the

conversion table.

YES, NO

NO

Menu
No.

Function

Function description

Selection list

Default