OPERATION
6
73
OPTIWAVE 5200 C/F
www.krohne.com
07/2013 - 4001904902 - HB OPTIWAVE 5200 R02 en
2.5.7
EMP.SPEC.EN.
The supervisor can start or stop the empty
spectrum filter.
YES, NO
NO
2.6.0 COMMUNICAT.
2.6.1
ADDRESS
Any HART
®
address greater than 0 will
activate HART
®
multidrop mode. The current
output stays constant at 4 mA.
min.-max: 0…63
0
2.7.0 DISPLAY
2.7.1
LANGUAGE
Data can be shown in any of the languages
stored in the device.
9 languages are
available in 3 packs:
(1) English, French,
German and Italian;
(2) English, French,
Spanish and
Portuguese;
(3) English, Chinese
(Mandarin), Japanese
and Russian
3
2.7.2
LENGTH UNIT
The length unit shown in normal mode.
m, cm, mm, in (inches),
ft (feet)
m
2.7.3
CONV UNIT
Conversion unit. The length, volume, mass or
flow rate conversion unit for the conversion
table and shown in normal mode.
m3, L, gal (US gallons),
ImpG (Imperial
gallons), ft3, bbl (oil
barrel), kg, t, Ston,
Lton, m, cm, mm, in, ft,
m3/h, ft3/h
L
2.7.4
PSWD YES/NO
If it is necessary to protect your settings in
the supervisor menu with a password, set
this menu item to YES
YES
YES
YES.
YES, NO
YES
2.7.5
PASSWORD
This changes the password for the supervisor
menu. Push the buttons up to 6 times in any
order. This will be the new password. To
confirm the change, enter the new password
a second time. For more data, refer to
Protection of the device settings
on page 78.
[>>>>], [^
^
^
^], [
],
[
], [>>>>] and
[^
^
^
^]
2.7.6
CONTRAST
The contrast control for the display screen.
You can select a shade of grey between light
grey (level 20) and black (level 54).
min.-max: 20…54
36
2.8.0 CONV. TABLE
2.8.1
INPUT TABLE
The device uses a conversion table (strapping
table) to convert measurements to volume,
mass and flow rate readings. The readings
are shown in normal mode. Give the number
of entries on the table (min. 2; max. 30).
Select an entry (01...30) and enter the level
and the related volume / mass / flow rate
value for that entry. Push [^
^
^
^] to confirm the
entry values. Continue the procedure until
the device has data for all the entries. For
more data, refer to
How to configure the
device to measure volume or mass
on page
81.
min. 2 entries
max. 30 entries
(level / volume, mass
or flow rate)
0 entries
2.8.2
DELETE TABLE
This menu item erases the data in the
conversion table.
YES, NO
NO
Menu
No.
Function
Function description
Selection list
Default