If an "Out of specification" or a "Maintenance" status symbol is shown, refer to menu item 2.2.2
DIAGNOSTIC (Configuration mode / Supervisor menu) for more data.

For data on errors, error records and error codes, refer to

Error handling

on page 86.

NE 107

symbol

shown

NE 107 Status

Description

Status

marker

shown

Error code

(Type)

Possible errors

Failure

The device does not operate

correctly. The fault message

stays on. The user cannot

remove the "Failure" message

from the Normal mode screen.

1

ERR 101 (Error)

Current Output Drift

3

ERR 102 (Error)

Temperature Out of Range

1

ERR 103 (Error)

Converter EEPROM

1

ERR 103 (Error)

Converter RAM

1

ERR 103

(Error)

Converter ROM

1

ERR 104 (Error)

Converter Voltage

1

ERR 200 (Error)

Internal Communication

2, 4

ERR 201 (Error)

Overfill

2, 4

ERR 203 (Error)

Peak Lost (Level Lost)

3

ERR 204 (Error)

Sensor no Signal

1

ERR 206 (Error)

Sensor Microwave

1

ERR 207 (Error)

Sensor EEPROM

1

ERR 207 (Error)

Sensor RAM

1

ERR 207 (Error)

Sensor ROM

1

ERR 208 (Error)

Sensor Voltage

1

ERR 210 (Error)

Sensor Not compatible

Function

check

The device operates correctly,

but the measured value is

incorrect. This fault message

is only temporary. This symbol

is shown when the user

configures the device with the

DTM or a HART®

Communicator.

-

-

-

Out of

specification

It is possible that the

measured value is unstable if

the operating conditions do not

agree with the device

specification.

4

(Warning)

Peak Lost

4

(Warning)

Overfill

3

(Warning)

Temperature out of range

Maintenance

The device does not operate

correctly because of bad

environmental conditions (e.g.

build-up on the antenna). The

measured value is correct, but

maintenance is neccessary a

short time after this symbol is

shown.

5

(Warning)

Empty Spectrum Invalid

4

(Warning)

Signal Weak

4

(Warning)

Signal Strong

4

(Warning)

Bad Measurement Quality

3

(Warning)

Temperature <-35°C / -31°F

3

(Warning)

Temperature >+75°C / +167°F