Technical data – KROHNE OPTIWAVE 5200 C_F EN User Manual
Page 105
TECHNICAL DATA
8
105
OPTIWAVE 5200 C/F
www.krohne.com
07/2013 - 4001904902 - HB OPTIWAVE 5200 R02 en
FOUNDATION
FOUNDATION
FOUNDATION
FOUNDATION™ fieldbus
fieldbus
fieldbus
fieldbus
Physical layer
FOUNDATION™ fieldbus protocol that agrees with IEC 61158-2 and FISCO model
Communication standard
H1
ITK version
6.1
Function blocks
1 × Resource Block (RB), 3 × Transducer Blocks (TB), 3 × Analog Input Blocks (AI), 1
x Proportional Integral Derivative Block (PID)
Analog Input Block: 30 ms
Proportional Integral Derivative Block: 40 ms
Device power supply
Not intrinsically safe: 9...32 VDC
Intrinsically safe: 9...24 VDC
Basic current
14 mA
Maximum error current
20.5 mA (= basic current + error current = 14 mA + 6.5 mA)
Polarity sensitivity
No
Minimum cycle time
250 ms
Output data
Level, distance, ullage conversion, level conversion
Input data
None
Error current FDE
Typically 0 mA (FDE =Fault Disconnection Electronic)
Link Master function
Supported
Approvals and certification
CE
This device fulfils the statutory requirements of the EC directives. The
manufacturer certifies successful testing of the product by applying the CE mark.
Vibration resistance
EN 60068-2-64
Metallic Horn (without antenna extension options): 5 Hz to 100 Hz: 4g
Metallic Horn, PTFE or PP Wave Horn: 3.5 mm up to 8 Hz and 10 m/s²: 1g, 8.5 to
2000 Hz
Explosion protection
Explosion protection
Explosion protection
Explosion protection
ATEX (Ex ia or Ex d)
DEKRA 11ATEX0166 X
Compact version
Compact version
Compact version
Compact version
II 1/2 G, 2 G Ex ia IIC T6...T2 Ga/Gb or Ex ia IIC T6...T2 Gb;
II 1/2 D, 2 D Ex ia IIIC T90°C Da/Db or Ex ia IIIC T90°C Db IP6X;
II 1/2 G, 2 G Ex d ia IIC T6...T2 Ga/Gb or Ex d ia IIC T6...T2 Gb;
II 1/2 D, 2 D Ex ia tb IIIC T90°C Da/Db or Ex ia tb IIIC T90°C Db IP6X
Remote version, transmitter
Remote version, transmitter
Remote version, transmitter
Remote version, transmitter
II 2 G Ex ia [ia Ga] IIC T6...T4 Gb;
II 2 D Ex ia [ia Da] IIIC T90°C Db;
II 2 G Ex d ia [ia Ga] IIC T6...T4 Gb;
II 2 D Ex ia tb [ia Da] IIIC T90°C Db
Remote version, sensor
Remote version, sensor
Remote version, sensor
Remote version, sensor
II 1/2 G Ex ia IIC T6...T2 Ga/Gb
II 1/2 D Ex ia IIIC T90°C Da/Db
II 1/2 G Ex ia IIC T6...T2 Gb
II 1/2 D Ex ia IIIC T90°C Db