beautypg.com

Configure gpib interface – GW Instek GDS-3000 Series User Manual User Manual

Page 183

background image

REMOTE CONTROL CONFIG

183

Configure GPIB Interface

To use GPIB, the optional GPIB to USB (GUG-001) adapter must be
used. The GPIB address can be configured for the GUG-001 from
the utility menu. See the GUG-001 user manual for more
information.

Configure GPIB

1. Insert the GUG-001 USB cable into

the rear panel USB device port.

2. Press the Utility key.

Utility

3. Press I/O from the bottom menu.

4. Press USB Device Port from the

side menu.

5. Press GPIB from the side menu.

6. Use the variable knob to set the

GPIB Address from the side menu.

Range

1 ~ 30

GPIB Constraints

Maximum 15 devices altogether, 20m cable
length, 2m between each device

Unique address assigned to each device

At least 2/3 of the devices turned On

No loop or parallel connection