beautypg.com

Teledyne LeCroy PXA125 User Manual

Page 214

background image

User Manual PXA125

Specifications A-96

Memory Interleave

4 (All trace lengths must be multiples of 4)

Vertical Resolution

14 bits (16,384 points)

Waveform Download Rate

5 Meg points per second

Sine Wave Performance

Description

Sine wave performance is measured using the maximum sample clock rate,

at 5 Vp-p, the maximum amplitude resolution of the DAC (14-bit) and without filters. Spectrum ana-
lyzer settings: Start frequency = 0 MHz; stop frequency = 50 MHz (Nyquist frequency)

Sinewave Total Harmonic Distortion

0.05% to 100 KHz

Harmonics and non-related spurious

< -55dBc for carrier frequencies

1MHz

< -40dBc for carrier frequencies

5MHz

< -35dBc for carrier frequencies

10MHz

< -22dBc for carrier frequencies

50MHz

SEQUENCED ARBITRARY WAVEFORMS

Operation

Permits division of the memory bank into smaller segments. Segments may

be linked, and repeated in user-selectable fashion to generate extremely long waveforms.

Advance Modes

Automatic Sequence Advance No triggers required to step from one segment to the

next. Sequence is repeated continuously through a pre-programmed sequence list (table)

Stepped Sequence Advance

Current segment is sampled continuously, external

trigger advances to next programmed segment. Control input is TRIG IN connector.

Single Sequence Advance

Current segment is sampled to the end of the seg-

ment including repeats and idles there. Next trigger advances to next segment. Control input is TRIG
IN connector.

Mixed Sequence Advance

Each step of a sequence can be programmed to ad-

vance either a) automatically (Automatic Sequence Advance), or b) with a trigger (Stepped Sequence
Advance)

Advance Source

External, Internal, or soft trigger

Sequencer steps

From 1 to 4096

Segment loops From 1 to 128 K