Output subsystem – Teledyne LeCroy PXA125 User Manual
Page 173
User Manual PXA125
Programming Reference 4-55
Parameter range
<%_amplitude> -100 to 100 in units of percent
DC?
Response and default
The PXA125 will return the present dc value. Default value is 100.
OUTPut
Subsystem
This subsystem controls the characteristics of the output. It controls
filter type and filter cutoff frequency. It is also used to control the sync
output, source and position. Optional nodes were omitted from these
commands. Factory defaults after *RST are shown in bold typeface.
Parameter low and high limits are given where applicable.
Keyword
Parameter Form
Default,Low
Limit,High
Limit
:OUTPut (?)
{OFF|ON|0|1}
OFF
:FILTer (?)
{NONE|25M|50M|ALL}
NONE
:SYNC (?)
{OFF|ON|0|1}
OFF
:SOURce (?)
{BIT|LCOMplete}
BIT
:POSition (?)
0,0,n
:WIDTh (?)
4,4,100000
:TTLTrig
{1|2|3|4|5|6|7}
[:STATe] (?)
{OFF|ON|0|1}
OFF
:STAR
{1|2|3|4|5|6|7}
[:STATe] (?)
{OFF|ON|0|1}
OFF
OUTPut{OFF|ON|0|1}
Purpose
This command will turn the PXA125 output on and off.
•
ON or 1 will set the output on
•
OFF or 0 will set the output off
Parameter type
Discrete
OUTPut?
Response and default
The PXA125 will return 1 if the output is on, or 0 if the output is off. Default value is 0.
OUTPut:FILTer{NONE|25M|50M|ALL}
Purpose
This command will select which filter is connected to the PXA125 output. Observe the following restrictions
when you try to use this command: