User Manual PXA125

Programming Reference 4-54

Parameter range

4 to 100 zero crossings

SINC:NCYCle?

Response and default
The PXA125 will return the present number of zero-crossing value. Default value is 10.

GAUSsian:EXPonent

Purpose
This command programs the exponent for the standard gaussian pulse waveform. This command has no
affect on arbitrary waveforms.

sets the exponent parameter

Parameter type

Numeric

Parameter range

10 to 200

GAUSsian:EXPonent?

Response and default
The PXA125 will return the present exponent value. Default value is 20.

EXPonential:EXPonent

Purpose
This command programs the exponent for the standard exponential waveform. This command has no affect
on arbitrary waveforms.

sets the exponent parameter

Parameter type

Numeric

Parameter range

-100 to 100

EXPonential:EXPonent?

Response and default
The PXA125 will return the present exponent value. Default value is –20.

DC<%_amplitude>

Purpose
This command programs the amplitude of the standard DC waveform. This command has no affect on
arbitrary waveforms.

<%_amplitude > sets the dc level as percentage of the programmed amplitude setting

Parameter type

Numeric