Programming Reference 4-57
Parameter type
Discrete
OUTPut:FILTer?
Response and default
The PXA125 will return NONE, 25M, 50M, or ALL depending on the type of filter presently connected to the
output. Default value is NONE.
OUTPut:SYNC{OFF|ON|0|1}
Purpose
This command will turn the PXA125 SYNC output on and off.
•
ON or 1 will set the SYNC output on
•
OFF or 0 will set the SYNC output off
Parameter type
Discrete
OUTPut:SYNC?
Response and default
The PXA125 will return 1 if the SYNC output is on, or 0 if the SYNC output is off. Default value is 0.
OUTPut:SYNC:SOURce{BIT|LCOMplete}
Purpose
This command will select the PXA125 SYNC option.
•
BIT will set the SYNC to output a narrow pulse
•
LCOMplete will set the SYNC to transition high at the beginning of the sequence and low a the end of the
sequence
Parameter type
Discrete
OUTPut:SYNC:SOURce?
Response and default
The PXA125 will return BIT, or LCOM depending on the present setting of the SYNC source. Default value is
BIT.
OUTPut:SYNC:POSition
Purpose
This command will program the PXA125 SYNC position. This command is active in arbitrary (USER) mode
only.
•
Parameter type
Numeric (integer only)