beautypg.com

Teledyne LeCroy PXA125 User Manual

Page 185

background image

User Manual PXA125

Programming Reference 4-67

TRIGger:BURSt{OFF|ON|0|1}

Purpose
This command will set the burst mode. This command will affect the PXA125 only after it will be set to
interrupted operation with the command INIT:CONT OFF.

ON or 1 will set the burst mode on

OFF or 0 will set the burst mode off

Parameter type

Discrete

TRIGger:BURSt?

Response and default
The PXA125 will return 1 if the burst is on, or 0 if the burst is off. Default value is 0.

TRIGger:COUNt

Purpose
This command will set the burst counter.

will set the count number

Parameter type

Numeric (integer only)

Parameter range

1 to 128K

TRIGger:COUNt?

Response and default
The PXA125 will return the present count value. Default value is 1.

TRIGger:GATE{OFF|ON|0|1}

Purpose
This command will set the gate mode. This command will affect the PXA125 only after it will be set to
interrupted operation with the command INIT:CONT OFF.

ON or 1 will set the gate mode on

OFF or 0 will set the gate mode off

Parameter type

Discrete

TRIGger:GATE?

Response and default