Teledyne LeCroy PXA125 User Manual
Page 185
User Manual PXA125
Programming Reference 4-67
TRIGger:BURSt{OFF|ON|0|1}
Purpose
This command will set the burst mode. This command will affect the PXA125 only after it will be set to
interrupted operation with the command INIT:CONT OFF.
•
ON or 1 will set the burst mode on
•
OFF or 0 will set the burst mode off
Parameter type
Discrete
TRIGger:BURSt?
Response and default
The PXA125 will return 1 if the burst is on, or 0 if the burst is off. Default value is 0.
TRIGger:COUNt
Purpose
This command will set the burst counter.
•
Parameter type
Numeric (integer only)
Parameter range
1 to 128K
TRIGger:COUNt?
Response and default
The PXA125 will return the present count value. Default value is 1.
TRIGger:GATE{OFF|ON|0|1}
Purpose
This command will set the gate mode. This command will affect the PXA125 only after it will be set to
interrupted operation with the command INIT:CONT OFF.
•
ON or 1 will set the gate mode on
•
OFF or 0 will set the gate mode off
Parameter type
Discrete
TRIGger:GATE?
Response and default