Teledyne LeCroy PXA125 User Manual
Page 164
User Manual PXA125
Programming Reference 4-46
ROSCillator:SOURce{INTernal|EXTernal|TCXO}
Purpose
This command will select the reference source for the sample clock generator.
•
•
PXA125 for it to continue normal operation
•
purchase the instrument with the TCXO option, this command will not be available for you to use.
Parameter type
Discrete
ROSCillator:SOURce?
Response and default
The PXA125 will return INT, EXT, or TCXO depending on the present PXA125 setting. Default value is INT.
SWEep:STOP
Purpose
This command will set the stop sample clock frequency. When set to operate in sweep mode, the PXA125
will sweep from carrier sample clock frequency to sweep stop sample clock frequency.
•
Parameter type
Numeric
Parameter range
100e-3 to 100e6 in units of samples per second
SWEep:STOP?
Response and default
The PXA125 will return the present sweep stop sample clock frequency value. The returned value will be in
standard scientific format (for example: 1KHz would be returned as 1E3 – positive numbers are unsigned).
Default value is 20e6.
SWEep:TIME
Purpose
This command will set the time it takes for the sample clock frequency to sweep from carrier to stop
frequencies. This parameter will affect the PXA125 only when the instrument is set to operate in sweep
mode.
•
Parameter type
Numeric
Parameter range