User Manual PXA125

Programming Reference 4-50

VOLTage?

Response to query version
The PXA125 will return the present amplitude value. The returned value will be in standard scientific format
(for example: 100mV would be returned as 100E-3 – positive numbers are unsigned). Default value is 5.

VOLTage:OFFSet

Purpose
This command programs the amplitude offset of the output waveform. The offset is calibrated when the
source impedance is 50

.

sets the amplitude offset

Parameter type

Numeric

Parameter range

-3.6 to +3.6 in units of volts

VOLTage:OFFSet?

Response and default
The PXA125 will return the present offset value. The returned value will be in standard scientific format (for
example: 100mV would be returned as 100E-3 – positive numbers are unsigned). Default value is 0.

SINusoid:PHASe

Purpose
This command programs start phase of the standard sine waveform. This command has no affect on arbitrary
waveforms.

sets the start phase

Parameter type

Numeric

Parameter range

0 to 360 in units of degrees

SINusoid:PHASe?

Response and default
The PXA125 will return the present start phase value. Default is 0.

TRIangle:PHASe

Purpose
This command programs start phase of the standard triangular waveform. This command has no affect on
arbitrary waveforms.

sets the start phase

Parameter type

Numeric