Teledyne LeCroy PXA125 User Manual
Page 157

User Manual PXA125
Programming Reference 4-39
APPLy:USER?
Response
The PXA125 will return
FM{OFF|ON|0|1}
Purpose
This command will turn the FM function on and off.
•
ON or 1 will set the FM on
•
OFF or 0 will set the FM off
Parameter type
Boolean
FM?
Response and default
The PXA125 will return 1 if the FM is on, or 0 if the FM is off. Default is 0.
FM:DEViation
Purpose
This command will set the deviation frequency in units of samples per second.
•
Parameter type
Numeric
Parameter range
1e-3 to 100e6 in units of samples per second
FM:DEViation?
Response and default
The PXA125 will return the present frequency deviation value. The returned value will be in standard scientific
format (for example: 100mHz would be returned as 100E-3 – positive numbers are unsigned). Default value
is 1e3. Default value is 1e6.
FM:FUNCtion:MODE{FIXed|USER}
Purpose
This command will set the frequency modulating waveform mode to FIXed (standard) and USER (arbitrary)
waveforms.
•
FIXed will enable a library of built in standard waveforms
•
USER will set the generator to accept arbitrary waveform from an external application
Parameter type
Discrete