Programming Reference 4-42

Numeric

Parameter range

100e-6 to 50e6

FREQuency?

Response and default
The PXA125 will return the present frequency value. The returned value will be in standard scientific format
(for example: 100mHz would be returned as 100E-3 – positive numbers are unsigned). Default value is 1e6.

FREQuency:RASTer{|MINimum|MAXimum}

Purpose
This command modifies the sample clock frequency of the arbitrary waveform in units of samples per second
(S/s). It has no affect on standard waveforms.

sets the sample clock frequency

sets the sample clock frequency to its lowest value

sets the sample clock frequency to its highest value

Parameter type

Numeric

Parameter range

100e-3 to 100e6

FREQuency:RASTer?

Response and default
The PXA125 will return the present sample clock frequency value. The returned value will be in standard
scientific format (for example: 100MHz would be returned as 100E6 – positive numbers are unsigned).
Default value is 10e6.

FREQuency:RASTer:SOURce{EXTernal|INTernal}

Purpose
This command selects the source of the sample clock generator. This command affects both the standard
and the arbitrary waveforms.

selects an external source

selects the internal source

Parameter type

Discrete

FREQuency:RASTer:SOURce?

Response and default
The PXA125 will return EXT if an external source is selected, or INT if the internal source is selected. Default
value is INT.