User Manual PXA125

Programming Reference 4-48

Response and default
The PXA125 will return CONT, TRIG, or GATE depending on its present setting. Default is CONT.

SWEep:TRIGger:SLOPe{POSitive|NEGative}

Purpose
This command will select slope sensitivity for the sweep trigger input.

POSitive will select rising edge

NEGative will select falling edge

Parameter type

Discrete

SWEep:TRIGger:SLOPe?

Response and default
The PXA125 will return POS, or NEG depending on its present setting. Default is POS.

SWEep:MARKer

Purpose
This command will set the marker sample clock frequency for the sweep mode.

will set the marker sample clock frequency

Parameter type

Numeric

Parameter range

100e-3 to 100e6 in units of samples per second

SWEep:MARKer?

Response and default
The PXA125 will return the present sweep sample clock frequency value. The returned value will be in
standard scientific format (for example: 1KHz would be returned as 1E3 – positive numbers are unsigned).
Default value is 64e6.

VOLTage{|MINimum|MAXimum}

Purpose
This command programs the peak-to-peak amplitude of the output waveform. The amplitude is calibrated
when the source impedance is 50

.

sets the amplitude

sets the amplitude to its lowest value

sets the amplitude to its highest value

Parameter type

Numeric

Parameter range

10e-3 to 10 in units of volts

10e-3 in units of volts