Teledyne LeCroy PXA125 User Manual
Page 184
User Manual PXA125
Programming Reference 4-66
ARM:SLOPe{POSitive|NEGative}
Purpose
This command will set the edge sensitivity of which the PXA125 will stop in arm mode. The stop and start
commands are applied at the rear-panel TRIG IN connector.
•
POSitive will set the positive edge
•
NEGative will set the negative edge
Parameter type
Discrete
ARM:SLOPe?
Response and default
The PXA125 will return POS, or NEG depending on the present setting. Default value is POS.
ARM:BREakpoint:POSition
Purpose
This command will program the breakpoint position for the arm function.
•
Parameter type
Numeric (integer only)
Parameter range
0 to “n” in units of waveform points, where “n” is the size of the active segment. The
breakpoint position can be programmed in increments of 4 points minimum.
ARM:BREakpoint:POSition?
Response and default
The PXA125 will return the present breakpoint value. Default value is 0.
INITiate:CONTinuous{OFF|ON|0|1}
Purpose
This command will set the output in continuous operation and interrupted operation. The trigger subsystem
commands will affect the PXA125 only after it will be set to interrupted operation.
•
ON or 1 will set the continuous mode
•
OFF or 0 will set the interrupted mode
Parameter type
Discrete
INITiate:CONTinuous?
Response to query version
The PXA125 will return 1 if the output is continuous, or 0 if the output is interrupted. Default value is 0.