2 technical specifications – iris 220 – Xylem SDI-12 Data Logger iRIS 220, iRIS 320, iRIS 350, iRIS 150, iRIS 300 User Manual
Page 81

75
iQuest (NZ) Ltd - PO Box 15169, Hamilton, New Zealand Tel: +64 7 857-0810 Fax: +64 7 857-0811 Email: [email protected]
12.2 Technical Specifications – iRIS 220
Dimensions:
100mm x 172mm x 40mm (3.93in x 6.77in x 1.57in)
(Excluding connectors)
Mass:
380g (13.4oz)
Power Supply:
External regulated 12V dc power supply or rechargeable sealed lead-acid battery.
Charger Source - External 15-30V dc supply. Supports a directly connected solar
panel (no regulator required).
Power Consumption:
Less than 3mA @ 12V in idle mode and full power save mode selected.
Actual current consumption is dependent on power management mode, wireless
modem state and I/O configuration.
Comms Interfaces:
1x RS232 DB9, 38400 bps, DTE configuration.
1x Integral dual-band (900/1800MHz) GSM/GPRS modem or
Integral single-band (1900MHz) GSM/GPRS modem or
Integral CDMA/CDMA-1X modem or
Integral HSDPA (3G) modem
Digital I/O:
2 x Digital Inputs
- 30Vdc maximum input, switch to 0Vdc to activate.
- Link selectable for either mechanical (<20Hz) or electronic (<5kHz) switching.
2 x Digital Outputs
- 1 x Open-drain pull-down (max 300mA @ 30V)
- 1 x Switched 12V out (max 300mA)
Analog I/O:
4 x 12 bit uni-polar analog inputs. Range 0-
5000mV. Input impedance 98kΩ.
Referenced to 0V common. Internal measurements available for monitoring are:
Battery Voltage
Supply Voltage
Internal Temperature (C or F)
Received Signal Strength Indication (RSSI)
Logging Memory:
Non-volatile 2MB flash storage of 262,144 time/date stamped data points. Circular
buffer mode (overwrites oldest data when memory is full).
A typical site with 2 parameters logged every 15 minutes plus battery voltage logged
hourly will give 3.3 years of storage before data overwrite occurs.
Clock/Calendar:
Non-volatile with integral lithium back-up battery. Leap year recognition. Accurate to
+/-20 secs month, trimmable via a configuration register.
Mounting:
4 x M4 (3/16”) mounting holes at ends of case.
Environmental:
Storage Temperature: -20C - +85C. (-4F - +185F)
Operating Temperature: -10C - +70C. (14F - +158F)